Quantification of LV function and mass by cardiovascular magnetic resonance : multi-center variability and consensus contours

  • Background: High reproducibility of LV mass and volume measurement from cine cardiovascular magnetic resonance (CMR) has been shown within single centers. However, the extent to which contours may vary from center to center, due to different training protocols, is unknown. We aimed to quantify sources of variation between many centers, and provide a multi-center consensus ground truth dataset for benchmarking automated processing tools and facilitating training for new readers in CMR analysis. Methods: Seven independent expert readers, representing seven experienced CMR core laboratories, analyzed fifteen cine CMR data sets in accordance with their standard operating protocols and SCMR guidelines. Consensus contours were generated for each image according to a statistical optimization scheme that maximized contour placement agreement between readers. Results: Reader-consensus agreement was better than inter-reader agreement (end-diastolic volume 14.7 ml vs 15.2–28.4 ml; end-systolic volume 13.2 ml vs 14.0–21.5 ml; LV mass 17.5 g vs 20.2–34.5 g; ejection fraction 4.2 % vs 4.6–7.5 %). Compared with consensus contours, readers were very consistent (small variability across cases within each reader), but bias varied between readers due to differences in contouring protocols at each center. Although larger contour differences were found at the apex and base, the main effect on volume was due to small but consistent differences in the position of the contours in all regions of the LV. Conclusions: A multi-center consensus dataset was established for the purposes of benchmarking and training. Achieving consensus on contour drawing protocol between centers before analysis, or bias correction after analysis, is required when collating multi-center results.

Download full text files

Export metadata

Additional Services

Share in Twitter Search Google Scholar
Metadaten
Author:Avan Suinesiaputra, David A. Bluemke, Brett R. Cowan, Matthias G. Friedrich, Christopher M. Kramer, Raymond Y. Kwong, Sven Plein, Jeanette Schulz-Menger, Jos J. M. Westenberg, Alistair A. Young, Eike NagelORCiDGND
URN:urn:nbn:de:hebis:30:3-335672
DOI:https://doi.org/10.1186/s12968-015-0170-9
ISSN:1532-429X
ISSN:1097-6647
Pubmed Id:https://pubmed.ncbi.nlm.nih.gov/26215273
Parent Title (English):Journal of cardiovascular magnetic resonance
Publisher:BioMed Central
Place of publication:London
Document Type:Article
Language:English
Date of Publication (online):2015/07/28
Date of first Publication:2015/07/28
Publishing Institution:Universitätsbibliothek Johann Christian Senckenberg
Release Date:2015/07/28
Tag:Benchmarks; Image analysis; Left ventricular function; Left ventricular mass; Multi-center; Reproducibility
Volume:17
Issue:63
Page Number:8
Note:
© 2015 Suinesiaputra et al. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
HeBIS-PPN:369699599
Institutes:Medizin / Medizin
Dewey Decimal Classification:6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Sammlungen:Universitätspublikationen
Licence (German):License LogoCreative Commons - Namensnennung 4.0