• Treffer 6 von 9
Zurück zur Trefferliste

Postoperative radiological assessment of the mastoid facial canal in cochlear implant patients in correlation with facial nerve stimulation

  • Objectives: To correlate the radiological assessment of the mastoid facial canal in postoperative cochlear implant (CI) cone-beam CT (CBCT) and other possible contributing clinical or implant-related factors with postoperative facial nerve stimulation (FNS) occurrence. Methods: Two experienced radiologists evaluated retrospectively 215 postoperative post-CI CBCT examinations. The mastoid facial canal diameter, wall thickness, distance between the electrode cable and mastoid facial canal, and facial-chorda tympani angle were assessed. Additionally, the intracochlear position and the insertion angle and depth of electrodes were evaluated. Clinical data were analyzed for postoperative FNS within 1.5-year follow-up, CI type, onset, and causes for hearing loss such as otosclerosis, meningitis, and history of previous ear surgeries. Postoperative FNS was correlated with the measurements and clinical data using logistic regression. Results: Within the study population (mean age: 56 ± 18 years), ten patients presented with FNS. The correlations between FNS and facial canal diameter (p = 0.09), wall thickness (p = 0.27), distance to CI cable (p = 0.44), and angle with chorda tympani (p = 0.75) were statistically non-significant. There were statistical significances for previous history of meningitis/encephalitis (p = 0.001), extracochlear-electrode-contacts (p = 0.002), scala-vestibuli position (p = 0.02), younger patients’ age (p = 0.03), lateral-wall-electrode type (p = 0.04), and early/childhood onset hearing loss (p = 0.04). Histories of meningitis/encephalitis and extracochlear-electrode-contacts were included in the first two steps of the multivariate logistic regression. Conclusion: The mastoid-facial canal radiological assessment and the positional relationship with the CI electrode provide no predictor of postoperative FNS. Histories of meningitis/encephalitis and extracochlear-electrode-contacts are important risk factors.

Volltext Dateien herunterladen

Metadaten exportieren

Metadaten
Verfasserangaben:Iris BurckORCiDGND, Rania A. Helal, Nagy Naguib Naeem NaguibORCiDGND, Nour-Eldin Abdelrehim Nour-Eldin MohammedORCiDGND, Jan-Erik ScholtzGND, Simon MartinORCiDGND, Martin LeinungORCiDGND, Silke HelbigORCiDGND, Timo StöverGND, Annette LehnORCiDGND, Thomas J. VoglORCiDGND
URN:urn:nbn:de:hebis:30:3-636079
DOI:https://doi.org/10.1007/s00330-021-08128-w
ISSN:1432-1084
Titel des übergeordneten Werkes (Englisch):European radiology
Verlag:Springer
Verlagsort:Berlin ; Heidelberg
Dokumentart:Wissenschaftlicher Artikel
Sprache:Englisch
Datum der Veröffentlichung (online):05.07.2021
Datum der Erstveröffentlichung:05.07.2021
Veröffentlichende Institution:Universitätsbibliothek Johann Christian Senckenberg
Datum der Freischaltung:01.06.2022
Freies Schlagwort / Tag:Cochlear implantation; Cone-beam computed tomography; Electric stimulation; Facial nerve
Jahrgang:32
Ausgabe / Heft:1
Seitenzahl:9
Erste Seite:234
Letzte Seite:242
Bemerkung:
Open Access funding enabled and organized by Projekt DEAL. The authors state that this work has not received any funding.
HeBIS-PPN:496362224
Institute:Medizin
DDC-Klassifikation:6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Sammlungen:Universitätspublikationen
Lizenz (Deutsch):License LogoCreative Commons - Namensnennung 4.0