Heat generation at the implant–bone interface by insertion of ceramic and titanium implants

  • Purpose: The aim of this study is to record material- and surface-dependent heat dissipation during the process of inserting implants into native animal bone. Materials and Methods: Implants made of titanium and zirconium that were identical in macrodesign were inserted under controlled conditions into a bovine rib tempered to 37 °C. The resulting surface temperature was measured on two bone windows by an infrared camera. The results of the six experimental groups, ceramic machined (1), sandblasted (2), and sandblasted and acid-etched surfaces (3) versus titanium implants with the corresponding surfaces (4, 5, and 6) were statistically tested. Results: The average temperature increase, 3 mm subcrestally at ceramic implants, differed with high statistical significance (p = 7.163 × 10−9, resulting from group-adjusted linear mixed-effects model) from titanium. The surface texture of ceramic implants shows a statistical difference between group 3 (15.44 ± 3.63 °C) and group 1 (19.94 ± 3.28 °C) or group 2 (19.39 ± 5.73 °C) surfaces. Within the titanium implants, the temperature changes were similar for all surfaces. Conclusion: Within the limits of an in vitro study, the high temperature rises at ceramic versus titanium implants should be limited by a very slow insertion velocity.

Download full text files

Export metadata

Metadaten
Author:Holger ZipprichGND, Paul WeiglORCiD, Eugenie König, Alexandra Toderas, Ümniye BalabanORCiDGND, Christoph Martin RatkaORCiDGND
URN:urn:nbn:de:hebis:30:3-512770
DOI:https://doi.org/10.3390/jcm8101541
ISSN:2077-0383
Pubmed Id:https://pubmed.ncbi.nlm.nih.gov/31557930
Parent Title (English):Journal of Clinical Medicine
Publisher:MDPI
Place of publication:Basel
Document Type:Article
Language:English
Year of Completion:2019
Date of first Publication:2019/09/25
Publishing Institution:Universitätsbibliothek Johann Christian Senckenberg
Release Date:2019/10/14
Tag:bone damage; bone–implant interface; dental implant; early loss; heat; insertion; zirconia
Volume:8
Issue:10, Art. 1541
Page Number:17
First Page:1
Last Page:17
Note:
This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited
HeBIS-PPN:455370524
Institutes:Medizin / Medizin
Dewey Decimal Classification:6 Technik, Medizin, angewandte Wissenschaften / 61 Medizin und Gesundheit / 610 Medizin und Gesundheit
Sammlungen:Universitätspublikationen
Open-Access-Publikationsfonds:Medizin
Licence (German):License LogoCreative Commons - Namensnennung 4.0