Dosing accuracy of two disposable insulin pens according to new ISO 11608-1: 2012 requirements.
(2015)
OBJECTIVE: The aim was to compare 2 disposable insulin pens, FlexTouch® (Novo Nordisk, insulin aspart) and SoloSTAR® (Sanofi, insulin glulisine), according to new ISO 11608-1:2012 requirements for dosing accuracy.
METHODS: Sixty pens of each type were tested at 1, 40, and 80 U doses. Following the new ISO requirements, each dose was delivered from the front, middle, and rear one-third of the pen. Statistical analysis was performed using Student's t test.
RESULTS: Both pens delivered all doses within ISO limits. The difference between the average measured dose and the target dose was significantly smaller for SoloSTAR than FlexTouch at 40 U (P = .009) and 80 U (P = .008), but not at 1 U (P = .417).
CONCLUSION: Both insulin pens fulfilled the dosing accuracy requirements defined by ISO 11608-1:2012 at all 3 dosage levels.
This paper reports on Monte Carlo simulation results for future measurements of the moduli of time-like proton electromagnetic form factors, |GE | and |GM|, using the ¯pp → μ+μ− reaction at PANDA (FAIR). The electromagnetic form factors are fundamental quantities parameterizing the electric and magnetic structure of hadrons. This work estimates the statistical and total accuracy with which the form factors can be measured at PANDA, using an analysis of simulated data within the PandaRoot software framework. The most crucial background channel is ¯pp → π+π−,due to the very similar behavior of muons and pions in the detector. The suppression factors are evaluated for this and all other relevant background channels at different values of antiproton beam momentum. The signal/background separation is based on a multivariate analysis, using the Boosted Decision Trees method. An expected background subtraction is included in this study, based on realistic angular distribuations of the background contribution. Systematic uncertainties are considered and the relative total uncertainties of the form factor measurements are presented.