Purpose: High precision radiosurgery demands comprehensive delivery-quality-assurance techniques. The use of a liquid-filled ion-chamber-array for robotic-radiosurgery delivery-quality-assurance was investigated and validated using several test scenarios and routine patient plans.
Methods and material: Preliminary evaluation consisted of beam profile validation and analysis of source–detector-distance and beam-incidence-angle response dependence. The delivery-quality-assurance analysis is performed in four steps: (1) Array-to-plan registration, (2) Evaluation with standard Gamma-Index criteria (local-dose-difference ⩽ 2%, distance-to-agreement ⩽ 2 mm, pass-rate ⩾ 90%), (3) Dose profile alignment and dose distribution shift until maximum pass-rate is found, and (4) Final evaluation with 1 mm distance-to-agreement criterion. Test scenarios consisted of intended phantom misalignments, dose miscalibrations, and undelivered Monitor Units. Preliminary method validation was performed on 55 clinical plans in five institutions.
Results: The 1000SRS profile measurements showed sufficient agreement compared with a microDiamond detector for all collimator sizes. The relative response changes can be up to 2.2% per 10 cm source–detector-distance change, but remains within 1% for the clinically relevant source–detector-distance range. Planned and measured dose under different beam-incidence-angles showed deviations below 1% for angles between 0° and 80°. Small-intended errors were detected by 1 mm distance-to-agreement criterion while 2 mm criteria failed to reveal some of these deviations. All analyzed delivery-quality-assurance clinical patient plans were within our tight tolerance criteria.
Conclusion: We demonstrated that a high-resolution liquid-filled ion-chamber-array can be suitable for robotic radiosurgery delivery-quality-assurance and that small errors can be detected with tight distance-to-agreement criterion. Further improvement may come from beam specific correction for incidence angle and source–detector-distance response.
BACKGROUND: We retrospectively evaluated the efficacy and toxicity of gross tumor volume (GTV) mean-dose-optimized and real-time motion-compensated robotic stereotactic body radiation therapy (SBRT) in the treatment of liver metastases.
METHODS: Between March 2011 and July 2015, 52 patients were treated with SBRT for a total of 91 liver metastases (one to four metastases per patient) with a median GTV volume of 12 cc (min 1 cc, max 372 cc). The optimization of mean GTV dose was prioritized during treatment planning at the potential cost of planning target volume (PTV) coverage reduction while adhering to safe normal tissue constraints. The delivered median GTV biological effective dose (BED10) was 142.1 Gy10 (range, 60.2 Gy10 -165.3 Gy10) and the prescribed PTV BED10 ranged from 40.6 Gy10 to 112.5 Gy10 (median, 86.1 Gy10). We analyzed local control (LC), progression-free interval (PFI), overall survival (OS), and toxicity.
RESULTS: Median follow-up was 17 months (range, 2-49 months). The 2-year actuarial LC, PFI, and OS rates were 82.1, 17.7, and 45.0 %, and the median PFI and OS were 9 and 23 months, respectively. In univariate analysis histology (p < 0.001), PTV prescription BED10 (HR 0.95, CI 0.91-0.98, p = 0.002) and GTV mean BED10 (HR 0.975, CI 0.954-0.996, p = 0.011) were predictive for LC. Multivariate analysis showed that only extrahepatic disease status at time of treatment was a significant factor (p = 0.033 and p = 0.009, respectively) for PFI and OS. Acute nausea or fatigue grade 1 was observed in 24.1 % of the patients and only 1 patient (1.9 %) had a side effect of grade ≥ 2.
CONCLUSIONS: Robotic real-time motion-compensated SBRT is a safe and effective treatment for one to four liver metastases. Reducing the PTV prescription dose and keeping a high mean GTV dose allowed the reduction of toxicity while maintaining a high local control probability for the treated lesions.