Refine
Year of publication
- 1992 (1)
Document Type
- Article (1)
Language
- English (1)
Has Fulltext
- yes (1)
Is part of the Bibliography
- no (1)
Keywords
Institute
- Medizin (1)
A comparison of different APTT-reagents, heparin-sensitivity and detection of mild coagulopathies
(1992)
The activated partial thromboplastin time (aPTT) is widely used to detect coagulation abnormalities or to monitor heparin treatment.
Many commercial aPTT-reagents are available which contain different phospholipid reagents and activators. In the present study 3 aPTT-reagents (aPTT-D, Instrumentation Laboratory, Neothromtin, Behring, PTTa, Boehringer) were compared using a computerized centrifugal analyzer. One aPTT-reagent (Pathromtin, Behring) was tested on a semiautomated coagulometer. Instrument precision was evaluated using aPTT-D as reagent.
Comparative tests were performed on plasma samples of 40 healthy donors, 3 patients with mild von Willebrand's disease (vWd), W patients with heaemophilia or subhaemophilia A, 1 patient with subhaemophilia A and vWd, 8 patients treated with subcutaneous injection of unfractionated heparin (UFH) and 14 patients treated with subcutaneous injection of a low molecular weight heparin (LMWH).
aPTT-D was the most sensitive reagent to detect mild vWd while Pathromtin detected none of these defects. In patients with heamophilia A and subhaemophilia A aPTT-D, Neothromtin and PTTa detected the abnormality in nearly all tested samples while Pathromtin was less sensitive.
Patients treated with subcutaneously applied UFH or LMWH often had a prolonged aPTTt especially when aPTT-D and Neothromtin were used as reagents.