Background aims: Immunomagnetic enrichment of CD34+ hematopoietic “stem” cells (HSCs) using paramagnetic nanobead coupled CD34 antibody and immunomagnetic extraction with the CliniMACS plus system is the standard approach to generating T-cell-depleted stem cell grafts. Their clinical beneficence in selected indications is established. Even though CD34+ selected grafts are typically given in the context of a severely immunosuppressive conditioning with anti-thymocyte globulin or similar, the degree of T-cell depletion appears to affect clinical outcomes and thus in addition to CD34 cell recovery, the degree of T-cell depletion critically describes process quality. An automatic immunomagnetic cell processing system, CliniMACS Prodigy, including a protocol for fully automatic CD34+ cell selection from apheresis products, was recently developed. We performed a formal process validation to support submission of the protocol for CE release, a prerequisite for clinical use of Prodigy CD34+ products.
Methods: Granulocyte-colony stimulating factor–mobilized healthy-donor apheresis products were subjected to CD34+ cell selection using Prodigy with clinical reagents and consumables and advanced beta versions of the CD34 selection software. Target and non-target cells were enumerated using sensitive flow cytometry platforms.
Results: Nine successful clinical-scale CD34+ cell selections were performed. Beyond setup, no operator intervention was required. Prodigy recovered 74 ± 13% of target cells with a viability of 99.9 ± 0.05%. Per 5 × 10E6 CD34+ cells, which we consider a per-kilogram dose of HSCs, products contained 17 ± 3 × 10E3 T cells and 78 ± 22 × 10E3 B cells.
Conclusions: The process for CD34 selection with Prodigy is robust and labor-saving but not time-saving. Compared with clinical CD34+ selected products concurrently generated with the predecessor technology, product properties, importantly including CD34+ cell recovery and T-cell contents, were not significantly different. The automatic system is suitable for routine clinical application.
Background: Because of limitations of transportation imposed by the COVID-19 pandemic, current recommendation calls for cryopreservation of allogeneic stem cell transplants before patient conditioning. A single cell therapy laboratory was selected to function as the central cryopreservation hub for all European registry donor transplants intended for the Australian-Pacific region. We examined properties of these transplants to ascertain how quality is maintained.
Methods: We analyzed 100 pandemic-related allogeneic mobilized blood-derived stem cell apheresis products generated at 30 collection sites throughout Europe, shipped to and cryopreserved at our center between April and November of 2020. Products were shipped in the cool, subsequently frozen with DMSO as cryoprotectant. Irrespective of origin, all products were frozen within the prescribed shelf-life of 72 h.
Results: Prior to cryopreservation, viable stem cell and leukocyte count according to the collection site and our reference laboratory were highly concordant (r2 = 0.96 and 0.93, respectively) and viability was > 90% in all instances. Median nominal post-thaw recovery of viable CD34+ cells was 42%. Weakly associated with poorer CD34+ cell recovery was higher leukocyte concentration, but not time lag between apheresis or addition of cryopreservant, respectively, and start of freezing. The correlation between pre- and post-thaw CD34+ cell dose was high (r2 = 0.85), hence predictable. Neutrophil and platelet engraftment were prompt with no evidence of dose dependency within the range of administered cell doses (1.31–15.56 × 106 CD34+ cells/kg).
Conclusions: General cryopreservation of allogeneic stem cell transplants is feasible. While more than half of the CD34+ cell content is lost, the remaining stem cells ensure timely engraftment.