Background. Tracheal intubation still represents the "gold standard" in securing the airway of unconscious patients in the prehospital setting. Especially in cases of restricted access to the patient, video laryngoscopy became more and more relevant.
Objectives. The aim of the study was to evaluate the performance and intubation success of four different video laryngoscopes, one optical laryngoscope, and a Macintosh blade while intubating from two different positions in a mannequin trial with difficult access to the patient.
Methods. A mannequin with a cervical collar was placed on the driver’s seat. Intubation was performed with six different laryngoscopes either through the driver’s window or from the backseat. Success, C/L score, time to best view (TTBV), time to intubation (TTI), and number of attempts were measured. All participants were asked to rate their favored device.
Results. Forty-two physicians participated. 100% of all intubations performed from the backseat were successful. Intubation success through the driver’s window was less successful. Only with the Airtraq® optical laryngoscope, 100% success was achieved. Best visualization (window C/L 2a; backseat C/L 2a) and shortest TTBV (window 4.7 s; backseat 4.1 s) were obtained when using the D-Blade video laryngoscope, but this was not associated with a higher success through the driver’s window. Fastest TTI was achieved through the window (14.2 s) when using the C-MAC video laryngoscope and from the backseat (7.3 s) when using a Macintosh blade.
Conclusions. Video laryngoscopy revealed better results in visualization but was not associated with a higher success. Success depended on the approach and familiarity with the device. We believe that video laryngoscopy is suitable for securing airways in trapped accident victims. The decision for an optimal device is complicated and should be based upon experience and regular training with the device.
Objective. Evaluation of C-MAC PM® in combination with a standard Macintosh blade size 3 in direct and indirect laryngoscopy and D-Blade® in indirect laryngoscopy in a simulated difficult airway. Primary outcome was defined as the best view of the glottic structures. Secondary endpoints were subjective evaluation and assessment of the intubation process.
Methods. Prospective monocentric, observational study on 48 adult patients without predictors for difficult laryngoscopy/tracheal intubation undergoing orthopedic surgery. Every participant preoperatively received a cervical collar to simulate a difficult airway. Direct and indirect laryngoscopy w/o the BURP maneuver with a standard Macintosh blade and indirect laryngoscopy w/o the BURP maneuver using D-Blade® were performed to evaluate if blade geometry and the BURP maneuver improve the glottic view as measured by the Cormack-Lehane score.
Results. Using a C-MAC PM® laryngoscope, D-Blade® yielded improved glottic views compared with the Macintosh blade used with either the direct or indirect technique. Changing from direct laryngoscopy using a Macintosh blade to indirect videolaryngoscopy using C-MAC PM® with D-Blade® improved the Cormack-Lehane score from IIb, III, or IV to I or II in 31 cases.
Conclusion. The combination of C-MAC PM® and D-Blade® significantly enhances the view of the glottis compared to direct laryngoscopy with a Macintosh blade in patients with a simulated difficult airway.
Trial Registration Number. This trial is registered under number NCT03403946.