Refine
Year of publication
- 2020 (2) (remove)
Document Type
- Article (2)
Language
- English (2)
Has Fulltext
- yes (2) (remove)
Is part of the Bibliography
- no (2) (remove)
Keywords
- Multiplate (1)
- NIRS (1)
- Quantra (1)
- TAVI (1)
- algorithm (1)
- aortic stenosis (1)
- bypass (1)
- cardiothoracic surgery (1)
- coagulopathy (1)
- delirium (1)
Institute
- Medizin (2)
Background: Point of care devices for performing targeted coagulation substitution in patients who are bleeding have become increasingly important in recent years. New on the market is the Quantra. It is a device that uses sonorheometry, a sonic estimation of elasticity via resonance, which is a novel ultrasound-based technology that measures viscoelastic properties of whole blood. Several studies have already shown the comparability of the Quantra with devices already established on the market, such as the rotational thromboelastometry (ROTEM) device.
Objective: In contrast to existing studies, this study is the first prospective interventional study using this new system in a cardiac surgical patient cohort. We will investigate the noninferiority between an already existing coagulation algorithm based on the ROTEM/Multiplate system and a new algorithm based on the Quantra system for the treatment of coagulopathic cardiac surgical patients.
Methods: The study is divided into two phases. In an initial observation phase, whole blood samples of 20 patients obtained at three defined time points (prior to surgery, after completion of cardiopulmonary bypass, and on arrival in the intensive care unit) will be analyzed using both the ROTEM/Multiplate and Quantra systems. The obtained threshold values will be used to develop a novel algorithm for hemotherapy. In a second intervention phase, the new algorithm will be tested for noninferiority against an algorithm used routinely for years in our department.
Results: The main objective of the examination is the cumulative loss of blood within 24 hours after surgery. Statistical calculations based on the literature and in-house data suggest that the new algorithm is not inferior if the difference in cumulative blood loss is <150 mL/24 hours.
Conclusions: Because of the comparability of the Quantra sonorheometry system with the ROTEM measurement methods, the existing hemotherapy treatment algorithm can be adapted to the Quantra device with proof of noninferiority.
Trial Registration: ClinicalTrials.gov NCT03902275; https://clinicaltrials.gov/ct2/show/NCT03902275
International Registered Report Identifier (IRRID): DERR1-10.2196/17206
Background: Cerebral O2 saturation (ScO2) reflects cerebral perfusion and can be measured noninvasively by near-infrared spectroscopy (NIRS). Objectives: In this pilot study, we describe the dynamics of ScO2 during TAVI in nonventilated patients and its impact on procedural outcome. Methods and Results: We measured ScO2 of both frontal lobes continuously by NIRS in 50 consecutive analgo-sedated patients undergoing transfemoral TAVI (female 58%, mean age 80.8 years). Compared to baseline ScO2 dropped significantly during RVP (59.3% vs. 53.9%, p < .01). Five minutes after RVP ScO2 values normalized (post RVP 62.6% vs. 53.9% during RVP, p < .01; pre 61.6% vs. post RVP 62.6%, p = .53). Patients with an intraprocedural pathological ScO2 decline of >20% (n = 13) had higher EuroSCORE II (3.42% vs. 5.7%, p = .020) and experienced more often delirium (24% vs. 62%, p = .015) and stroke (0% vs. 23%, p < .01) after TAVI. Multivariable logistic regression revealed higher age and large ScO2 drops as independent risk factors for delirium. Conclusions: During RVP ScO2 significantly declined compared to baseline. A ScO2 decline of >20% is associated with a higher incidence of delirium and stroke and a valid cut-off value to screen for these complications. NIRS measurement during TAVI procedure may be an easy to implement diagnostic tool to detect patients at high risks for cerebrovascular complications and delirium.