OBJECTIVES: Outcome of aortic valve replacement may be influenced by the choice of bioprosthesis. Pericardial heart valves are described to have a favourable haemodynamic profile compared with porcine valves, although the clinical notability of this finding is still controversially debated. Herein, we compared the long-term results of two commonly implanted bioprosthesis at a single centre.
METHODS: All consecutive patients undergoing isolated aortic valve replacement with either a Carpentier-Edwards Magna pericardial prosthesis or a Medtronic Mosaic porcine prosthesis between 2002 and 2008 were analysed regarding preoperative characteristics, short- and long-term survival, valve-related complications and echocardiographic findings.
RESULTS: The Medtronic Mosaic was implanted in 163 patients and the Carpentier-Edwards Magna in 295 patients. The sizes of implanted valves were 22.4 ± 1.5 mm for the Mosaic and 21.8 ± 1.8 mm for the Magna (P = 0.001). The long-term survival rate was 76 and 56% after 5 and 10 years for the Medtronic Mosaic, which was comparable with the Carpentier-Edwards Magna (77 and 57%; P = 0.92). Overall long-term survival was comparable with an age- and sex-matched Austrian general population for both groups. Valve-related adverse events were similar between groups. The postoperative mean transvalvular gradient was significantly increased in the Mosaic group (24 ± 9 mmHg vs 17 ± 7 mmHg; P < 0.001).
CONCLUSIONS: Both types of aortic bioprostheses offer excellent results after isolated aortic valve replacement. Despite relevant differences in gradients, long-term survival was comparable with the expected normal survival for both bioprostheses. Patients with a porcine heart valve had a higher postoperative transvalvular gradient.
Background: The aim of this study was to compare outcome of patients with previous cardiac surgery undergoing transapical aortic valve implantation (Redo-TAVI) to those undergoing classic aortic valve replacement (Redo-AVR) by using propensity analysis.
Methods: From January 2005 through May 2012, 52 high-risk patients underwent Redo-TAVI using a pericardial xenograft fixed within a stainless steel, balloon-expandable stent (Edwards SAPIEN™). During the same period of time 167 patients underwent classic Redo-AVR. Logistic regression analysis was used to identify covariates among 11 baseline patient variables including the type of initial surgery. Using the significant regression coefficients, each patient’s propensity score was calculated, allowing selectively matched subgroups of 40 patients each. Initial surgery included coronary artery bypass grafting in 30 patients, aortic valve replacement in 7 patients and mitral valve reconstruction in 3 patients in each group. Follow-up was 4 ± 2 years and was 100% complete.
Results: Postoperative chest tube drainage (163 ± 214 vs. 562 ± 332 ml/24 h, p = 0.02) and incidence of early permanent neurologic deficit (0 vs. 13%, p = 0.04) was lower in patients with Redo-TAVI and there was a trend towards improved 30-day survival (p = 0.06). Also we detected a decreased ventilation time (p = 0.04) and lower transfusion rate of allogenic blood products (p ≤ 0.05) in the Redo-TAVI group. At late follow up differences regarding incidence of major adverse events, including death and permanent neurologic deficits (25% vs. 43%, p = 0.01) statistically supported early postoperative findings.
Conclusion: The encouraging results regarding early and long-term outcomes following TAVI in patients with previous cardiac surgery show, that this evolving approach may be particularly beneficial in this patient cohort.