Refine
Year of publication
- 2016 (2) (remove)
Document Type
- Article (2)
Language
- English (2)
Has Fulltext
- yes (2)
Is part of the Bibliography
- no (2)
Keywords
- Automated Tube Potential Selection (1)
- Cancer (1)
- Cancer Staging (1)
- Chondral Lesion (1)
- Dual-Source CT (1)
- MEDIC (1)
- MRI (1)
- Multidetector Computed Tomography (1)
- Neoplasms (1)
- Patellofemoral Joint (1)
Institute
- Medizin (2)
BACKGROUND: Evaluation of latest generation automated attenuation-based tube potential selection (ATPS) impact on image quality and radiation dose in contrast-enhanced chest-abdomen-pelvis computed tomography examinations for gynaecologic cancer staging.
METHODS: This IRB approved single-centre, observer-blinded retrospective study with a waiver for informed consent included a total of 100 patients with contrast-enhanced chest-abdomen-pelvis CT for gynaecologic cancer staging. All patients were examined with activated ATPS for adaption of tube voltage to body habitus. 50 patients were scanned on a third-generation dual-source CT (DSCT), and another 50 patients on a second-generation DSCT. Predefined image quality setting remained stable between both groups at 120 kV and a current of 210 Reference mAs. Subjective image quality assessment was performed by two blinded readers independently. Attenuation and image noise were measured in several anatomic structures. Signal-to-noise ratio (SNR) was calculated. For the evaluation of radiation exposure, CT dose index (CTDIvol) values were compared.
RESULTS: Diagnostic image quality was obtained in all patients. The median CTDIvol (6.1 mGy, range 3.9-22 mGy) was 40 % lower when using the algorithm compared with the previous ATCM protocol (median 10.2 mGy · cm, range 5.8-22.8 mGy). A reduction in potential to 90 kV occurred in 19 cases, a reduction to 100 kV in 23 patients and a reduction to 110 kV in 3 patients of our experimental cohort. These patients received significantly lower radiation exposure compared to the former used protocol.
CONCLUSION: Latest generation automated ATPS on third-generation DSCT provides good diagnostic image quality in chest-abdomen-pelvis CT while average radiation dose is reduced by 40 % compared to former ATPS protocol on second-generation DSCT.
Purpose: To determine the value of the 2D multiple-echo data image combination (MEDIC) sequence relative to the short-tau inversion recovery (STIR) sequence regarding the depiction of chondral lesions in the patellofemoral joint.
Materials and methods: During a period of 6 month patients with acute pain at the anterior aspect of the knee, joint effusion and suspected chondral lesion defect in the patellofemoral joint underwent MRI including axial MEDIC and STIR imaging. Patients with chondral lesions in the patellofemoral joint on at least one sequence were included. The MEDIC and STIR sequence were quantitatively compared regarding the patella cartilage-to-effusion contrast-to-noise ratio (CNR) and qualitatively regarding the depiction of chondral lesions independently scored by two radiologists on a 3-point scale (1 = not depicted; 2 = blurred depicted; 3 = clearly depicted) using the Wilcoxon-Mann-Whitney-Test. For the analysis of inter-observer agreement the Cohen's Weighted Kappa test was used.
Results: 30 of 58 patients (male: female, 21:9; age: 44 ± 12 yrs) revealed cartilage lesions (fissures, n = 5 including fibrillation; gaps, n = 15; delamination, n = 7; osteoarthritis, n = 3) and were included in this study. The STIR-sequence was significantly (p < 0.001) superior to the MEDIC-sequence regarding both, the patella cartilage-to-effusion CNR (mean CNR: 232 ± 61 vs. 40 ± 16) as well as the depiction of chondral lesion (mean score: 2.83 ± 0.4 vs. 1.75 ± 0.7) with substantial inter-observer agreement in the rating of both sequences (κ = 0.76–0.89).
Conclusion: For the depiction of chondral lesions in the patellofemoral joint, the axial STIR-sequence should be chosen in preference to the axial MEDIC-sequence.