Objectives: Combined electric-acoustic stimulation (EAS) is a well-accepted therapeutic treatment for cochlear implant (CI) users with residual hearing in the low frequencies but severe to profound hearing loss in the high frequencies. The recently introduced SONNETeas audio processor offers different microphone directionality (MD) settings and wind noise reduction (WNR) as front-end processing. The aim of this study was to compare speech perception in quiet and noise between two EAS audio processors DUET 2 and SONNETeas, to assess the impact of MD and WNR on speech perception in EAS users in the absence of wind. Furthermore, subjective rating of hearing performance was registered.
Method: Speech perception and subjective rating with SONNETeas or DUET 2 audio processor were assessed in 10 experienced EAS users. Speech perception was measured in quiet and in a diffuse noise setup (MSNF). The SONNETeas processor was tested with three MD settings omnidirectional/natural/adaptive and with different intensities of WNR. Subjective rating of auditory benefit and sound quality was rated using two questionnaires.
Results: There was no significant difference between DUET 2 and SONNETeas processor using the omnidirectional microphone in quiet and in noise. There was a significant improvement in SRT with MD settings natural (2.2 dB) and adaptive (3.6 dB). No detrimental effect of the WNR algorithm on speech perception was found in the absence of wind. Sound quality was rated as “moderate” for both audio processors.
Conclusions: The different MD settings of the SONNETeas can provide EAS users with better speech perception compared to an omnidirectional microphone. Concerning speech perception in quiet and quality of life, the performance of the DUET 2 and SONNETeas audio processors was comparable.
Introduction: Epoxyeicosatrienoic acids (EETs) are able to enhance angiogenesis and regulate inflammation that is especially important in wound healing under ischemic conditions. Thus, we evaluated the effect of local EET application on ischemic wounds in mice.
Methods: Ischemia was induced by cautherization of two of the three supplying vessels to the mouse ear. Wounding was performed on the ear three days later. Wounds were treated either with 11,12 or 14,15 EET and compared to untreated control and normal wounds. Epithelialization was measured every second day. VEGF, TNF-α, TGF-β, matrix metalloproteinases (MMP), tissue inhibitors of metalloproteinases (TIMP), Ki67, and SDF-1α were evaluated immunohistochemically in wounds on day 3, 6, and 9.
Results: Ischemia delayed wound closure (12.8 days ± 1.9 standard deviation (SD) for ischemia and 8.0 days ± 0.94 SD for control). 11,12 and14,15 EET application ameliorated deteriorated wound healing on ischemic ears (7.6 ± 1.3 SD for 11,12 EET and 9.2 ± 1.4 SD for 14,15 EET). Ischemia did not change VEGF, TNF-α, TGF-β, SDF-1α, TIMP, MMP7 or MMP9 level significantly compared to control. Local application of 11,12 as well as 14,15 EET induced a significant elevation of VEGF, TGF-β, and SDF-1α expression as well as proliferation during the whole phase of wound healing compared to control and ischemia alone.
Conclusion: In summary, EET improve impaired wound healing caused by ischemia as they enhance neovascularization and alter inflammatory response in wounds. Thus elevating lipid mediator level as 11,12 and 14,15 EET in wounds might be a successful strategy for amelioration of deranged wound healing under ischemia.